Variable Voltage Source (Opt. 309)
An internal DC source, available with Opt. 309, may be used to provide power
for external devices. The maximum power available is 8 watts.
The voltage is output through an SMB (m) output connector located on the
left-side panel. See the connector in
An optional SMB (f) to BNC (m) bias-tee power cable is available with Keysight
part number: N9910X-713.
The following status line is provided near the top of the FieldFox screen.
Figure 25-2
How to make Voltage Source settings
— Press
System 7
— Then
System Configuration
.
— Then
Voltage Source
.
— Then
Source ON OFF
to enable and disable the voltage source output.
— Then Voltage x.xx V Using
arrows, the rotary knob, or numeric
keypad, enter a value between 0.001 and 32 volts.
IMPORTANT!
Noise Figure mode uses DC variable voltage source (VVS) –
(Option 309) during the measurements. The voltage might be as high as
28V during the measurement. During the noise figure measurement,
always make sure there are no voltage sensitive devices connected to the
DC SMB output port. Refer to
Chapter 23, “Noise Figure Mode (Option 356
.
When you initially change the FieldFox mode to Noise Figure, the FieldFox
displays the following message for ten seconds:
"
Ensure your Noise Source is connected to the VVS DC Output
It (VVS) will be driven up to +28V during the measurement
".
Voltage Source Status Line
1. Voltage source state: ON, OFF, TRIPPED.
2.
Voltage
setting from 1.00 V to 32.00 V with.01 V resolution.
3.
Current:
The amount of current (in Amperes) being drawn from the source.
4.
Max Current: A
conservative estimate of maximum current possible given
the present voltage setting. The actual ‘trip’ level may be slightly different.
Содержание N9913A
Страница 6: ...6 ...
Страница 56: ...56 Keysight N9927 90001 User s Guide CAT Cable and Antenna Test Mode 2 Port Insertion Loss Measurements ...
Страница 82: ...82 Keysight N9927 90001 User s Guide TDR Time Domain Reflectometry Option 215 TDR Measurement Settings ...
Страница 106: ...106 Keysight N9927 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 142: ...142 Keysight N9927 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 202: ...202 Keysight N9927 90001 User s Guide SA Spectrum Analyzer Mode Channel Measurements ...
Страница 262: ...262 Keysight N9927 90001 User s Guide IQ Analyzer IQA Option 351 CPU2 Only Other Features Available ...
Страница 286: ...286 Keysight N9927 90001 User s Guide Interference Analyzer SA Mode Option 236 Record Playback SA Option ...
Страница 292: ...292 Keysight N9927 90001 User s Guide Reflection Mode SA Models Option 320 Reflection Mode Settings ...
Страница 316: ...316 Keysight N9927 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 332: ...332 Keysight N9927 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 346: ...346 Keysight N9927 90001 User s Guide VVM Vector Voltmeter Mode Option 308 A B and B A Measurements ...
Страница 370: ...370 Keysight N9927 90001 User s Guide AM FM Metrics SA Only and Option 355 Interpreting AM FM Metrics Error Messages ...
Страница 444: ...444 Keysight N9927 90001 User s Guide Noise Figure Mode Option 356 CPU2 Only Troubleshooting ...
Страница 470: ...Data Analysis Features All about Scale Settings 470 Keysight N9927 90001 User s Guide ...
Страница 528: ...528 Keysight N9927 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 544: ...544 Keysight N9927 90001 User s Guide Safety Considerations Certification and Compliance Statements ...
Страница 546: ...546 Keysight N9927 90001 User s Guide Connector Care Review ...
Страница 548: ...548 Keysight N9927 90001 User s Guide Connector Care Review ...
Страница 550: ...550 Keysight N9927 90001 User s Guide Specifications Data Sheet ...
Страница 553: ...553 Keysight Handheld Analyzers N99xxA User s Guide D Korean Battery Safety Translation ...
Страница 554: ...554 Keysight N9927 90001 User s Guide Korean Battery Safety Translation ...
Страница 555: ...555 Keysight Handheld Analyzers N99xxA User s Guide E Keysight Software End User Licensing Agreement EULA ...
Страница 556: ...556 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Страница 557: ...Keysight N9927 90001 User s Guide 557 Keysight Software End User Licensing Agreement EULA ...
Страница 558: ...558 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Страница 565: ......