80 Keysight N9927-90001 User’s Guide
TDR (Time Domain Reflectometry) - Option 215
TDR Measurement Settings
Start Frequency
– Start frequency that is used to calculate TDR.
Stop Frequency
– Stop frequency that is used to calculate TDR.
Range Resolution
. Indicates the accuracy of the distance to fault
measurement. For example, with range resolution of 500 mm, if the distance to
fault is 10 meters, this value could be inaccurate by +/- 500 mm or between
9.5 to 10.5 meters. This value is calculated from frequency span / resolution
(points).
Response Resolution, not displayed, indicates the distance that could be
between two faults and still show as separate faults. Learn more in Window
Maximum Distance
. The distance that could be viewed with the current
settings. Defined by: Vf*c*Points/(2*Bandwidth) where:
— Vf = velocity factor
— c = speed of light
— Points = resolution
— Bandwidth = frequency range
Alias-free Range (On/Off)
— On = No Alias images
— Off = Alias images may appear in the response.
About Alias Faults
An alias fault is not a true device response. An alias fault appears because of
the method used to convert frequency to time.
On the TDR Settings page (above) the
c - Alias-free Range = Off
setting
indicates alias images MAY appear on the screen.
Shorter stop distances (less than 10 meters) and a higher resolution (1001
points) will be more likely to result in
Alias-free Range = Off
.
When the Alias-free Range = Off, the following procedure will help to
determine if a response is true or an alias response:
1.
Put a marker on the response in question and note the distance to the
fault.
2.
Change the start or stop distance.
A true fault response will not move in distance. That is, if a true fault is present
at 10.3 meters, changing the stop distance from 15 m to 20 m will not move
the fault; the fault will remain at 10.3 meters. However, an alias response will
appear to move.
Содержание N9913A
Страница 6: ...6 ...
Страница 56: ...56 Keysight N9927 90001 User s Guide CAT Cable and Antenna Test Mode 2 Port Insertion Loss Measurements ...
Страница 82: ...82 Keysight N9927 90001 User s Guide TDR Time Domain Reflectometry Option 215 TDR Measurement Settings ...
Страница 106: ...106 Keysight N9927 90001 User s Guide NA Network Analyzer Mode NA Mode Settings ...
Страница 142: ...142 Keysight N9927 90001 User s Guide Calibration for NA CAT and VVM Modes Calibration Method Summary ...
Страница 202: ...202 Keysight N9927 90001 User s Guide SA Spectrum Analyzer Mode Channel Measurements ...
Страница 262: ...262 Keysight N9927 90001 User s Guide IQ Analyzer IQA Option 351 CPU2 Only Other Features Available ...
Страница 286: ...286 Keysight N9927 90001 User s Guide Interference Analyzer SA Mode Option 236 Record Playback SA Option ...
Страница 292: ...292 Keysight N9927 90001 User s Guide Reflection Mode SA Models Option 320 Reflection Mode Settings ...
Страница 316: ...316 Keysight N9927 90001 User s Guide Frequency Offset Using Power Sensor FOPS Option 208 FOPS Settings ...
Страница 332: ...332 Keysight N9927 90001 User s Guide Pulse Measurements Mode Option 330 Pulse Measurement Settings ...
Страница 346: ...346 Keysight N9927 90001 User s Guide VVM Vector Voltmeter Mode Option 308 A B and B A Measurements ...
Страница 370: ...370 Keysight N9927 90001 User s Guide AM FM Metrics SA Only and Option 355 Interpreting AM FM Metrics Error Messages ...
Страница 444: ...444 Keysight N9927 90001 User s Guide Noise Figure Mode Option 356 CPU2 Only Troubleshooting ...
Страница 470: ...Data Analysis Features All about Scale Settings 470 Keysight N9927 90001 User s Guide ...
Страница 528: ...528 Keysight N9927 90001 User s Guide Working with the Lithium Ion Battery Lithium Ion Battery Disposal ...
Страница 544: ...544 Keysight N9927 90001 User s Guide Safety Considerations Certification and Compliance Statements ...
Страница 546: ...546 Keysight N9927 90001 User s Guide Connector Care Review ...
Страница 548: ...548 Keysight N9927 90001 User s Guide Connector Care Review ...
Страница 550: ...550 Keysight N9927 90001 User s Guide Specifications Data Sheet ...
Страница 553: ...553 Keysight Handheld Analyzers N99xxA User s Guide D Korean Battery Safety Translation ...
Страница 554: ...554 Keysight N9927 90001 User s Guide Korean Battery Safety Translation ...
Страница 555: ...555 Keysight Handheld Analyzers N99xxA User s Guide E Keysight Software End User Licensing Agreement EULA ...
Страница 556: ...556 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Страница 557: ...Keysight N9927 90001 User s Guide 557 Keysight Software End User Licensing Agreement EULA ...
Страница 558: ...558 Keysight N9927 90001 User s Guide Keysight Software End User Licensing Agreement EULA ...
Страница 565: ......