Chapter 2
49
Test Descriptions
Frequency Accuracy
Performing the frequency accuracy calculations using the CSV
file in a spreadsheet
The following procedure uses Microsoft Excel.
Step 1. Open Excel.
Step 2. Open the .csv file from the A: drive. From the
File
menu select
Open
and
Look in:
(A:). Select files of
Type:
Text Files ( *.prn; *.txt; *.csv).
Figure 2-10
Example Excel File
Column A contains the measured frequency. Column B contains the
corresponding Linear P
hot
value. Disregard Column C null terminator
information.
Step 3. Increase the column widths to 20 to view the data more accurately.
Step 4. Enter the following formula into cell D1 = SUM(B$1:B1), copy this
formula and paste it from cell D2:D201.
Step 5. Enter the following formula into cell D202 =D201/2. This value returns
the ½ Power reading.
Содержание N8972A
Страница 9: ...ix Declaration of Conformity N8972A ...
Страница 10: ...x Declaration of Conformity N8973A ...
Страница 11: ...xi Declaration of Conformity N8974A ...
Страница 12: ...xii Declaration of Conformity N8975A ...
Страница 19: ...Contents xix ...
Страница 20: ...xx Contents ...
Страница 31: ...11 2 Test Descriptions This chapter descibes each test required for calibration and performance verification ...
Страница 108: ...88 Chapter2 Test Descriptions Measurement Jitter ...
Страница 124: ...104 Chapter3 Technical Specifications General Specifications ...
Страница 136: ...116 AppendixA Model N8972A Test Records Measurement Jitter Test Record ...
Страница 148: ...128 AppendixB Model N8973A Test Records Measurement Jitter Test Record ...
Страница 160: ...140 AppendixC Model N8974A Test Records Measurement Jitter Test Record ...
Страница 174: ...154 AppendixD Model N8975A Test Records Measurement Jitter Test Record ...