10 MHz Out Frequency Reference Adjustment
10 MHz Out Frequency Reference Adjustment
Test Description
The adjustment applies to NFAs with both the standard 10 MHz
frequency reference and 10 MHz precision frequency reference (Option
1D5).
The adjustment is performed by calculating the actual frequency error in
Hertz from Counter Reading 1 given in the 10 MHz Reference Accuracy
Test (see page 12). The negative or positive error is used to define what
the new DAC setting should be. After the DAC settings are adjusted and
saved, the 10 MHz Reference Accuracy Test is re-run to verify the unit
meets it’s published specification.
Required Test Equipment
The frequency standard provides the reference for the Universal
Counter. You can use a company-specific standard in place of the 5071A,
provided it meets the critical specifications listed in Table 2-3.
Table 2-3
10 MHz Out Frequency Reference Adjustment Test Equipment
Equipment description
Critical specification for equipment
substitution
Recommended model
Universal Counter
Time Interval Range: 25ms to 100ms
Single Operation Range: +2.5 to -2.5Vdc
53132A
Frequency Standard
Frequency: 10 MHz Timebase Accuracy
(Aging): < 1 x 10
-10
/day
5071A
BNC Lead (X 2)
Length 122cm (48 in.), Frequency
10 MHz
10503A
Содержание N8972A
Страница 9: ...ix Declaration of Conformity N8972A ...
Страница 10: ...x Declaration of Conformity N8973A ...
Страница 11: ...xi Declaration of Conformity N8974A ...
Страница 12: ...xii Declaration of Conformity N8975A ...
Страница 19: ...Contents xix ...
Страница 20: ...xx Contents ...
Страница 31: ...11 2 Test Descriptions This chapter descibes each test required for calibration and performance verification ...
Страница 108: ...88 Chapter2 Test Descriptions Measurement Jitter ...
Страница 124: ...104 Chapter3 Technical Specifications General Specifications ...
Страница 136: ...116 AppendixA Model N8972A Test Records Measurement Jitter Test Record ...
Страница 148: ...128 AppendixB Model N8973A Test Records Measurement Jitter Test Record ...
Страница 160: ...140 AppendixC Model N8974A Test Records Measurement Jitter Test Record ...
Страница 174: ...154 AppendixD Model N8975A Test Records Measurement Jitter Test Record ...