Model N8975A: Test Records
Noise Figure Range and Accuracy Test Record
Noise Figure Range and Accuracy Test Record
Results: 4.5 - 6.5dB Noise Source ENR
Input
Level
(dBm)
Range
(dB)
Actual
NF (dB)
Measured
NF (dB)
Noise Figure
Measurement
Range (dB)
Instrument
Uncertainty
(dB)
Specification
(dB)
Pass/
Fail
-----
0.125
0 - 20 dB
± 0.05dB
-----
0.5
± 0.05dB
-55
1
± 0.05dB
-56
2
± 0.05dB
-57
3
± 0.05dB
-58
4
± 0.05dB
-59
5
± 0.05dB
-60
6
± 0.05dB