19
Test Descriptions
10 MHz Out Frequency Reference Adjustment
Test Setup
Before starting the 10 MHz Out Frequency Reference Adjustment,
connect the equipment as shown in Figure 2-2.
Figure 2-2
10 MHz Out Frequency Reference Adjustment Test Setup
Test Procedure
NOTE
Before performing any adjustments ensure that the 10 MHz Frequency
Reference Accuracy test has been done and that the measured frequency
is outside the published specification.
NOTE
The NFA must be on and in internal frequency mode for at least 1 hour
before you start the test.
NOISE FIGURE
ANALYZER
FREQUENCY
STANDARD
UNIVERSAL
COUNTER
BNC Cable
10 MHz
Ref Out
Ref In
10 MHz
BNC Cable
Channel 1
Содержание N8972A
Страница 9: ...ix Declaration of Conformity N8972A ...
Страница 10: ...x Declaration of Conformity N8973A ...
Страница 11: ...xi Declaration of Conformity N8974A ...
Страница 12: ...xii Declaration of Conformity N8975A ...
Страница 19: ...Contents xix ...
Страница 20: ...xx Contents ...
Страница 31: ...11 2 Test Descriptions This chapter descibes each test required for calibration and performance verification ...
Страница 108: ...88 Chapter2 Test Descriptions Measurement Jitter ...
Страница 124: ...104 Chapter3 Technical Specifications General Specifications ...
Страница 136: ...116 AppendixA Model N8972A Test Records Measurement Jitter Test Record ...
Страница 148: ...128 AppendixB Model N8973A Test Records Measurement Jitter Test Record ...
Страница 160: ...140 AppendixC Model N8974A Test Records Measurement Jitter Test Record ...
Страница 174: ...154 AppendixD Model N8975A Test Records Measurement Jitter Test Record ...