Keysight M9514A Site Preparation and Installation Guide
23
Mains Disconnect
Connect the Chassis to the AC Power Line
Safety ground connection
This chassis must be grounded properly for proper operation and safety. Because
the M9514A is connected to the AC mains by a plug/socket connection, a
permanent earth ground must be supplied to reduce the risk of electric shock.
Make a permanent connection from the system rack to protective earth ground.
This connection serves as a redundant Protective Earth Connection to the
primary Protective Earth connection, which is part of the AC power cord. Refer to
. The “Earth (ground) terminal” found at the bottom of a system rack
should be connected by a wire separate from the AC power cord to the Protective
Earth connection at the AC source where the system AC power cord is
connected. The wire must be the same wire size as the protective conductor of
the system AC power cord. The wire may be either a bare conductor or a green
with yellow stripe insulated conductor.
The redundant Protective Earth connection wire should have a correctly sized
wire lug on both ends. The wire lugs should also be sized to fit the “Earth
(ground) terminal” stud or bolt as found on the system rack and for the
Protective Earth connection at the source end of system AC power cord. In some
cases, the Protective Earth connection at the source end of the system AC power
cord will receive the conductor without a wire lug.
Figure 1
Safety Ground Connection -- wire is not included with the chassis
System
Earth (ground)
Toothed washers
Wire Lug
Toothed
Bare wire or Green wire with
Wire Lug
Protective Earth
connection at the AC
source where the
The Safety Ground wire is not supplied with the M9514A chassis.
Содержание M9514A
Страница 1: ...Keysight AXIe 14 Slot Chassis M9514A SitePreparationand Installation Guide...
Страница 2: ......
Страница 6: ...vi...
Страница 14: ...16 Keysight M9514A Site Preparation and Installation Guide Introduction Plan the Chassis Installation...
Страница 37: ...Keysight M9514A Site Preparation and Installation Guide 39 Service and Support Additional Information...
Страница 40: ...42 Keysight M9514A AXIe 14 Slot Chassis Safety Related Specifications...
Страница 44: ...46 Keysight M9514A Site Preparation and Installation Guide Power Cords...