background image

M9514A AXIe Chassis 14-Slot

Site Preparation and Installation Guide

11

1

Introduction

Thank you for purchasing Agilent's M9514A AXIe 14-slot Chassis. To get you 

started and to assure a successful and timely installation, we’ve developed this 

Site Preparation Guide. Please read it carefully and follow all guidelines.
Correct site preparation is the key first step in ensuring that your chassis and 

software system operate reliably over an extended lifetime. This document is an 

information guide that outlines the utility requirements, space, and supplies for 

your chassis for your site.
After preparing the site and installing the M9514A chassis, review the:

-

M9514A AXIe Chassis Startup Guide. 

It provides chassis verification, installing 

AXIe instrument modules, connecting to a remote host computer or using an 

embedded controller. 

-

M9521A AXIe System Module Startup Guide. 

Provides installation and startup 

information about the M9521A ASM

-

M9514A and M9521A User Guide.

 It provides detailed information about the 

chassis, troubleshooting and service information, etc. 

Both of these guides are available on the on the 

Software and Product 

Information CD

 that ships with your chassis.

 

Customer Responsibilities

Make certain your site meets the following requirements prior to the installation. 

Details on these requirements are provided in this document.

If you have questions or problems in providing anything described in the 

Customer Responsibilities checklist above, please notify your local Agilent 

support/service organization for assistance prior to delivery. 

Operating environment (cooling, noise, etc.) is suitable. (

page 12

 and 

page 40

)

Receiving and transportation requirements are met (including uncrating). (

page 17

)

The necessary bench top or rack space is available. (

page 12

 and 

page 17

)

AC power requirements, power cords, etc. are met. (

page 12

 an

page 21

). 

Customer provides power cord (from mains disconnect to chassis) and IEC 60309 

(3P+N+E, 6H Red) receptacle.

Additional supplies (host computer, PCIe adapter, cables, LAN, etc.) are available. 

Содержание M9514A

Страница 1: ...Keysight AXIe 14 Slot Chassis M9514A SitePreparationand Installation Guide...

Страница 2: ......

Страница 3: ...IS SUBJECT TO BEING CHANGED WITHOUT NOTICE IN FUTURE EDI TIONS FURTHER TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW KEYSIGHT DISCLAIMS ALL WAR RANTIES EITHER EXPRESS OR IMPLIED WITH REGARD TO TH...

Страница 4: ...en with the equipment switched off To avoid dangerous elec trical shock DO NOT perform proce dures involving cover or shield removal unless you are qualified to do so DO NOT operate damaged equipment...

Страница 5: ...quipment WEEE Directive 2002 96 EC This product complies with the WEEE Directive 2002 96 EC marking requirement The affixed product label see below indicates that you must not discard this electrical...

Страница 6: ...vi...

Страница 7: ...he chassis 18 3 Connect the Chassis to the AC Power Line Mains Disconnect 22 Safety ground connection 23 Connecting Power to the Chassis 24 120 VAC Low Line 26 4 Rack Mount the Chassis General Rackmou...

Страница 8: ...iii Keysight M9514A Site Preparation and Installation Guide Y1237A 3 phase Delta Wye no Neutral 250 VAC NEMA L15 30P 2 5M 8 2ft 45 Y1238A 3 phase Wye with Neutral 240 415 VAC Stripped End 2 5M 8 2ft 4...

Страница 9: ...ion about the M9521A ASM M9514A and M9521A User Guide It provides detailed information about the chassis troubleshooting and service information etc Both of these guides are available on the on the So...

Страница 10: ...ining an M9514A AXIe chassis You should allow one meter of space in front of the rack for operator access and one meter of space behind the system for service Chassis Dimensions and Weight Identify th...

Страница 11: ...s the rack heat extractor fan noise should be considered Ambient noise levels in the room may be reduced with Dropped ceiling covered with a commercial grade of acoustic fiberglass flame resistant cei...

Страница 12: ...e cabling and associated hardware external to the chassis Refer to the M9514A and M9521A User Guide for additional information This product is designed for use in Installation Category II Pollution De...

Страница 13: ...mat and wrist strap combination Conductive floor mat and heel strap combination Both types when used together provide a significant level of ESD protection Of the two only the table mat and wrist str...

Страница 14: ...16 Keysight M9514A Site Preparation and Installation Guide Introduction Plan the Chassis Installation...

Страница 15: ...eceived all pieces of the system Your shipment should include the following M9514A Chassis1 does not include slot covers A box inside the shipping container with M9514A Site Preparation Guide printed...

Страница 16: ...ontain the power cord if you ordered one the printed documentation this Site Preparation Guide and the M9514A Startup Guide 4 Lift off the shipping carton 5 Use a mechanical lift to lift the M9514A an...

Страница 17: ...ument module slots Bottom front fan trays left to right Fan Tray 1 Fan Tray 2 AXIe System Module ASM Removable air filter On Stand by button ESD Wrist Strap Terminal Power supply status Chassis status...

Страница 18: ...ector refer to Connecting Power to the Chassis on page 24 for CAUTION For power supply fan tray removal only Do not lift chassis by AC Line Circuit Warning Do not lift chassis by these handles Chassis...

Страница 19: ...or 3 phase Wye with Neutral 240 415VAC Stripped End 2 5 M 8 2 feet 1 Refer to Appendix A Power Cords starting on page 43 for detailed information about the Agilent power cords An IEC 60309 3P N E 6H R...

Страница 20: ...d in the Off position but not in the On position Be installed within three meters of the chassis where it can be easily reached by the system operator without requiring the system to be moved to acces...

Страница 21: ...ection at the AC source where the system AC power cord is connected The wire must be the same wire size as the protective conductor of the system AC power cord The wire may be either a bare conductor...

Страница 22: ...ith Neutral 240 415VAC IEC 60309 2 5 M 8 2 feet 1 Y1237A for 3 phase Delta Wye no Neutral 250VAC NEMA L15 30P 2 5 M 8 2 feet 1 Y1238A for 3 phase Wye with Neutral 240 415VAC Stripped End 2 5 M 8 2 fee...

Страница 23: ...connector Agilent power cords have female connector shown on left Male connector on right is on M9514A rear panel N Earth L3 L2 L1 The mains wiring and connectors shall be compatible with the connect...

Страница 24: ...power up the ASM The chassis does not power up any other modules The chassis Soft Front Panel SFP will be able to connect to the chassis The M9514A ASM SFP or chassis web page will be able to connect...

Страница 25: ...paration and Installation Guide 27 120 VAC Low Line Connect the Chassis to the AC Power Line Figure 3 Normal Line voltage shown on the M9514A Web Page shows normal AC Line voltage PSU1 AC Input Voltag...

Страница 26: ...lation Guide Connect the Chassis to the AC Power Line 120 VAC Low Line The following graphic shows the sensors reading 120VAC Notice that the voltage lines are now red indicating Lower Non Recoverable...

Страница 27: ...support the chassis weight Do not use the support rails supplied with the rack 3 Always begin installing the chassis and heavy instruments at the bottom of the rack and work up The heaviest instrumen...

Страница 28: ...tip stabilizers and or ballast are recommended 9 To maintain adequate airflow do not install front doors on the rack If you have been using the chassis with modules installed you should remove all mo...

Страница 29: ...ps screwdriver to remove the rubber feet Store the feet and screws in case you should ever need to replace them 3 Carefully set the chassis upright again 4 Refer to Figure 6 Position the rack mounting...

Страница 30: ...a template for placing the clip on sheet metal nuts Mark the position the flange mounting holes on the rack s vertical columns where the chassis is to be mounted Slide clip on sheet metal nuts over t...

Страница 31: ...and Installation Guide 33 Procedure Rack Mount the Chassis Figure 7 Rackmounting the Agilent M9514A AXIe Chassis Clip on sheet metal nuts 4 required for rack Flange mounting holes 4 125 7 50in 5 25in...

Страница 32: ...hanges in airflow which can adversely affect cooling within the rack Cover these gaps with rack blank filler panels Chassis exceeds 48 kg Use a mechanical lift to lift the chassis The chassis should b...

Страница 33: ...is installed in a rack 1 Start by installing the support rails and chassis one EIA unit up from the bottom of the rack This prevents the rack base cover from interfering with the chassis See Figure 8...

Страница 34: ...are installed in slots in the rack always secure the rails to the rack 3 Install the second chassis on the support rails 4 Likewise to mount the third chassis the top chassis install the support rail...

Страница 35: ...the host PC Installing Agilent IO Libraries Suite and other chassis related software on the host PC Additionally the Software and Product Information CD that ships with your chassis has a PDF file of...

Страница 36: ...wn menu at the top Under Electronic Test and Measurement click on Services The Web page that appears next has contact information specific for your country Agilent by phone If you do not have access t...

Страница 37: ...Keysight M9514A Site Preparation and Installation Guide 39 Service and Support Additional Information...

Страница 38: ...characteristics AC input Operating voltage range Single Phase or 3 Phase Delta 3 Phase Wye 200 240 VAC 200 415 VAC Input frequency range 50 60 Hz Input current 24 Arms Overcurrent protection auto rec...

Страница 39: ...n type tested in accordance with the Keysight Environmental Test Manual and verified to be robust against the environmental stresses of storage transportation and end use those stresses include but ar...

Страница 40: ...42 Keysight M9514A AXIe 14 Slot Chassis Safety Related Specifications...

Страница 41: ...8 2ft The Agilent Y1235A Y1236A or Y1238A power cords are specifically designed to be used with the M9514A chassis you do not need to rewire the chassis power terminal block However if you are connect...

Страница 42: ...Guide Power Cords Y1236A 3 phase Wye with Neutral 240 415 VAC IEC 60309 2 5M 8 2ft Figure 11 Y1236A Wire Color L1 Brown L1 L2 Black L2 L3 Grey L3 N Blue N GN D Green Yellow GN D Splash proof IP44 IEC...

Страница 43: ...AC NEMA L15 30P 2 5M 8 2ft Y1238A 3 phase Wye with Neutral 240 415 VAC Stripped End 2 5M 8 2ft Figure 12 Y1237A Wire Color L1 Brown X L2 Black Y L3 Grey Z GN D Green Yellow G N Blue No Connectio n Fig...

Страница 44: ...46 Keysight M9514A Site Preparation and Installation Guide Power Cords...

Страница 45: ...the LXI consortium www pxisa org PCI eXtensions for Instrumentation PXI modular instrumentation delivers a rugged PC based high performance measurement and automation platform www keysight com find T...

Страница 46: ...This information is subject to change without notice Keysight Technologies 2014 Edition 2 September 2014 M9514 90007 M9514 90007 www keysight com...

Отзывы: