67
Keysight Infiniium 9000 Series Oscilloscopes
A Using Trigger Holdoff to Synchronize
Acquisition/Display on Complex Signals
Triggering on simple repetitive signals is easy using standard edge triggering.
But if you need to synchronize your oscilloscope’s acquisitions/display on more
complex signals, such as an amplitude-modulated signal, you will need to use
your oscilloscope’s trigger hold-off capability unless there is an external
synchronization signal available. This lab shows you how to use trigger holdoff
to achieve a stable trigger in the absence of a synchronization signal.
Make connections and perform initial setup:
1
To power the demo board, connect a USB cable between it and the
Keysight Infiniium 9000 Series oscilloscope.
You can also apply power by connecting the oscilloscope’s MSO cable to
the demo board’s 40-pin connector.
2
Connect the scope channel 1 probe to demo board CH1 and GND.
3 Make sure all other probes are disconnected from the oscilloscope
.
4
Turn the SELECT switch to position 1 (AM).
5
Press
[Default Setup]
.