Keysight Infiniium 9000 Series Oscilloscopes
4 InfiniiScan Triggering
Capturing infrequent or hard to describe anomalies such as random glitches
requires oscilloscopes with innovative technology. InfiniiScan is a
software-based post-processing technology that makes it easy to isolate
anomalies.
This lab requires an InfiniiScan option license.
Make connections and perform initial setup:
1
To power the demo board, connect a USB cable between it and the
Keysight Infiniium 9000 Series oscilloscope.
You can also apply power by connecting the oscilloscope’s MSO cable to
the demo board’s 40-pin connector.
2
Connect the scope channel 1 probe to demo board CH1 and GND.
3
Make sure all other probes are disconnected from the oscilloscope.
4
Turn the SELECT switch to position 3 (Burst).
5
Press
[Default Setup]
.
6
Press
[AutoScale]
.
7
Set the time base to 1 µs/div.
Set up the trigger:
8
From the on-screen main menu, choose
Trigger> InfiniiScan...
(or press
[Menu]
in the front panel Trigger controls).
9
In the Trigger Setup dialog, select:
Содержание Infiniium 9000 Series