49
Keysight Infiniium 9000 Series Oscilloscopes
8 SPI
To enable SPI serial decode, your oscilloscope has to have the I
2
C/SPI option
installed. (The installed options are listed in the
Help>About Infiniium...
dialog.)
Make connections and perform initial setup:
1
Connect the MSO cable between the oscilloscope and the demo board’s
40-pin connector.
This powers the demo board and probes its digital signals.
Normally, you would use flying leads to probe digital signals in your device
under test (DUT). The demo board 40-pin connector was designed with
built-in terminations.
2 Disconnect all oscilloscope probes
from the demo board.
This is a digital
signal only demo
.
3
Turn the SELECT switch to position 6 (SPI).
4
Press
[Default Setup]
.
5
Press
[AutoScale]
.
Set up serial decode:
6
Choose
Setup>Serial Decode...
.