3
Front-Panel Operation
84
Keysight E3633A and E3634A User’s Guide
HPIB Interface Configuration
The HPIB connector on the rear panel connects your power supply to the
computer and other HPIB devices. Chapter 1 lists the cables that are available
from Hewlett-Packard. An HPIB system can be connected together in any
configuration (star, linear, or both) as long as the following rules are observed:
– The total number of devices including the computer is no more than 15.
– The total length of all the cables used is no more than 2 meter times the
number of devices connected together, up to a maximum of 20 meters.
Do not stack more than three connector blocks together on any HPIB connector.
Make sure that all connectors are fully seated and that the lock screws are firmly
finger tightened.
NOTE
IEEE-488 states that you should exercise caution if your individual cable lengths
exceed 4 meters.
Содержание E3633A
Страница 1: ...Keysight E3633A and E3634A DC Power Supplies User s Guide ...
Страница 10: ...10 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 18: ...18 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 20: ...20 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 133: ...Remote Interface Reference 4 Keysight E3633A and E3634A User s Guide 133 SCPI Status System ...
Страница 166: ...5 Error Messages 166 Keysight E3633A and E3634A User s Guide 631 System DAC test failed 632 Hardware test failed ...
Страница 170: ...5 Error Messages 170 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 204: ...8 Specifications 204 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...