Remote Interface Reference
4
Keysight E3633A and E3634A User’s Guide
121
– To ensure synchronization when the bus source is selected, send the
*WAI
(wait) command. When the
*WAI
command is executed, the power supply
waits for all pending operations to complete before executing any additional
commands. For example, the following command string guarantees that the
first trigger is accepted and is executed before the second trigger is
recognized.
– You can use the
*OPC?
(operation complete query) command or the
*OPC
(operation complete) command to signal when the operation is complete. The
*OPC?
command returns “1” to the output buffer when the operation is
complete. The
*OPC
command sets the “OPC” bit (bit 0) in the Standard Event
register when the operation is complete.
Immediate triggering
– To select the immediate trigger source, send the following command.
– When the
IMMediate
is selected as a trigger source, an
INITiate
command
immediately transfers the
VOLT:TRIG
or
CURR:TRIG
value to
VOLT
or
CURR
value.
Any delay is ignored.
TRIG:SOUR BUS;*TRG;*WAI;*TRG;*WAI
TRIG:SOUR IMM
Содержание E3633A
Страница 1: ...Keysight E3633A and E3634A DC Power Supplies User s Guide ...
Страница 10: ...10 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 18: ...18 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 20: ...20 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 133: ...Remote Interface Reference 4 Keysight E3633A and E3634A User s Guide 133 SCPI Status System ...
Страница 166: ...5 Error Messages 166 Keysight E3633A and E3634A User s Guide 631 System DAC test failed 632 Hardware test failed ...
Страница 170: ...5 Error Messages 170 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 204: ...8 Specifications 204 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...