3
Front-Panel Operation
52
Keysight E3633A and E3634A User’s Guide
8 Verify that the power supply is in the constant voltage mode.
If you operate the power supply in the constant voltage (CV) mode, verify
that the
CV
annunciator is lit. If the
CC
annunciator is lit, choose a higher
current limit.
–
Remote interface operation:
NOTE
During actual CV operation, if a load change causes the current limit to be
exceeded, the power supply will automatically crossover to the constant current
mode at the preset current limit and the output voltage will drop
proportionately.
CURRent {<current>|MIN|MAX}
Set the current
VOLTage {<voltage>|MIN|MAX}
Set the voltage
OUTput ON
Enable the output
NOTE
You can use the resolution selection keys to move the blinking digit to the right
or left when setting voltage.
Содержание E3633A
Страница 1: ...Keysight E3633A and E3634A DC Power Supplies User s Guide ...
Страница 10: ...10 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 18: ...18 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 20: ...20 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 133: ...Remote Interface Reference 4 Keysight E3633A and E3634A User s Guide 133 SCPI Status System ...
Страница 166: ...5 Error Messages 166 Keysight E3633A and E3634A User s Guide 631 System DAC test failed 632 Hardware test failed ...
Страница 170: ...5 Error Messages 170 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 204: ...8 Specifications 204 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...