4
Remote Interface Reference
134
Keysight E3633A and E3634A User’s Guide
The questionable status register
The Questionable Status register provides information about voltage and current
regulation. Bit 0 is set when the voltage becomes unregulated, and bit 1 is set if
the current becomes unregulated. For example if the power supply momentarily
goes to constant current mode when the power supply is operating as a voltage
source (constant voltage mode), bit 0 is set to indicate that the voltage output is
not regulated.
The Questionable Status register also provides information that the power supply
has an overtemperature condition and that the overvoltage and overcurrent
protection circuits have tripped. Bit 4 reports an overtemperature condition of the
fan, bit 9 reports that the overvoltage protection circuit has tripped, and bit 10
reports that the overcurrent protection circuit has tripped. To read the register,
send
STATus:QUEStionable?
.
Table 4-3
Bit Definitions - Questionable Status Register
The questionable status event register is cleared when:
– You execute the
*CLS
(clear status) command.
– You query the event register using
STAT:QUES?
(Status Questionable Event
register) command.
For example, 16 is returned when you have queried the status of the questionable
event register, the temperature condition is questionable.
The questionable status enable register is cleared when:
– You execute
STAT:QUES:ENAB 0
command.
Bit
Decimal Value
Definition
0
Voltage
1
The power supply is/was in the constant current mode.
1
Current
2
The power supply is/was in the constant voltage mode.
2-3
Not Used
0
Always set to 0.
4
Overtemperature 16
The fan has a fault condition.
5-8
Not Used
0
Always set to 0.
9
Over Voltage
512
The overvoltage protection circuit has tripped.
10
Over Current
1024
The overvoltage protection circuit has tripped.
11-15 Not Used
0
Always set to 0.
Содержание E3633A
Страница 1: ...Keysight E3633A and E3634A DC Power Supplies User s Guide ...
Страница 10: ...10 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 18: ...18 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 20: ...20 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 133: ...Remote Interface Reference 4 Keysight E3633A and E3634A User s Guide 133 SCPI Status System ...
Страница 166: ...5 Error Messages 166 Keysight E3633A and E3634A User s Guide 631 System DAC test failed 632 Hardware test failed ...
Страница 170: ...5 Error Messages 170 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 204: ...8 Specifications 204 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...