CURRent:PROTection {<current>|MINimum|MAXimum}
This command sets the current level at which the overcurrent protection (OCP)
circuit will trip. If the peak output current exceeds the OCP level, then the
output current is programmed to zero. The Questionable Status register “OC”
bit is set (see
). An overcurrent condition can be cleared with the
CURR:PROT:CLE
command after the condition that caused the OCP trip is
removed.
CURRent:PROTection? {MINimum|MAXimum}
This query returns the overcurrent protection trip level presently programmed.
CURR:PROT? MAX
and
CURR:PROT?
MIN
return the maximum and minimum
programmable overcurrent trip levels.
CURRent:PROTection:STATe {0|1|OFF|ON}
This command enables or disables the overcurrent protection function of the
power supply. An overcurrent condition can be cleared with the
CURR:PROT:CLE
command after the condition that caused the OCP trip is removed. At
*RST
, this
value Is set to “ON”.
CURRent:PROTection:STATe?
This query returns the state of the overcurrent protection function. The
returned parameter is “0” (OFF) or “1” (ON).
CURRent:PROTection:TRIPped?
This query returns a “1” if the overcurrent protection circuit is tripped and not
cleared or a “0” if not tripped.
CURRent:PROTection:CLEar
This command causes the overcurrent protection circuit to be cleared. After
this command, the output current is restored to the state it was in before the
current protection tripped and the OCP trip level remains unchanged to the
value presently programmed. Before sending this command, lower the output
current below the trip OCP point, or raise the OCP trip level above the output
setting.
Note that the overcurrent condition caused by an external source must be
removed first before proceeding this command.
Содержание E3633A
Страница 1: ...Keysight E3633A and E3634A DC Power Supplies User s Guide ...
Страница 10: ...10 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 18: ...18 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 20: ...20 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 133: ...Remote Interface Reference 4 Keysight E3633A and E3634A User s Guide 133 SCPI Status System ...
Страница 166: ...5 Error Messages 166 Keysight E3633A and E3634A User s Guide 631 System DAC test failed 632 Hardware test failed ...
Страница 170: ...5 Error Messages 170 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 204: ...8 Specifications 204 Keysight E3633A and E3634A User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...