Transient response time
Less than 50
μ
sec for output to recover to within 15 mV following a change in
output current from full load to half load or vice versa
Command processing time
Average time for output voltage to begin to change after receipt of digital data
when the power supply is connected directly to the GPIB or RS-232 is less than
100 msec
OVP and OCP accuracy, ±(% of offset)
Activation time
Average time for output to start to drop after OVP and OCP condition occurs.
OVP
0.5% + 0.5 V
OCP
0.5% + 0.5 A
OVP
<1.5 msec when the trip voltage is equal or greater than 3 V
<10 msec when the trip voltage is less than 3 V
OCP
<10 msec