26
Getting Started Guide
1
Quick-Start
Ventilation
Instrument Maintenance
Cleaning the instrument
Cleaning the connectors
Battery Information
The test set uses a lithium battery located on the CPU board. This is not an operator
replaceable part. See
“Returning a test set for Service"
on page 113. Replaceable parts
must be approved or supplied by Agilent Technologies.
You can order the service documentation for the instrument through your Agilent Sales
and Service office.
VENTILATION REQUIREMENTS: When installing the product in a
cabinet, the convection into and out of the product must not be restricted.
The ambient temperature (outside the cabinet) must be less than the
maximum operating temperature of the product by 4
o
C for every 100
watts dissipated in the cabinet. If the total power dissipated in the
cabinet is greater than 800 watts, then forced convection must be used.
To prevent electrical shock, disconnect the test set from mains before
cleaning. Use a dry cloth or one slightly dampened with water to
clean the external case parts. Do not attempt to clean internally.
Cleaning connectors with alcohol shall only be done with the
instrument power cord removed, and in a well-ventilated area. Allow
all residual alcohol moisture to evaporate, and the fumers to
dissapate prior to energizing the instrument.
Danger of explosion if battery is incorrectly replaced. Replace only
with the same or equivalent type recommended. Discard used
batteries according to the manufacturer’s instructions.
Do not throw batteries away but collect as small chemical waste.
Содержание Agilent X-Series
Страница 9: ...10 Getting Started Guide...
Страница 17: ...18 Getting Started Guide 1 Quick Start The 1 GHz signal appears on the display Steps Actions Notes...
Страница 26: ...Quick Start 1 Getting Started Guide 27...
Страница 39: ...40 Getting Started Guide 2 Front and Rear Panel Features...
Страница 107: ...108 Getting Started Guide 4 Using Microsoft Windows Operating System...