Quick-Start
1
Getting Started Guide
17
The 50 MHz reference signal appears on the display.
8
Generate and view signal
a
Connect the ends of a short
RF cable to the RF Input
and RF Output connectors
on the front panel.
b
Press
Input/Output
,
RF Input
,
RF Input Port
,
RF In
.
c
Press
Input/Output
,
More
,
RF Output Port
,
RF Output
.
d
Press
Source
,
RF Output
,
On
.
e
Press
Source
,
Amplitude
,
RF Power
,
0
,
dBm
.
f
Press
Freq
,
Center Freq
,
1
,
GHz
.
g
Press
AMPTD Y Scale
,
Ref Value
,
10
,
dBm
.
h
Press
FREQ
,
Center Freq
,
1
,
GHz
.
i
Press
SPAN X Scale
,
Span
,
8
,
MHz
.
•
The purpose is to use a loopback cable to verify that
the test set’s source can generate a signal and the test
set’s analyzer can measure it.
•
Because there are four RF ports on the front panel, it
is necessary to specify which ports are used for the
input and output.
•
You need to choose frequency and power-level
settings for the source; you also need to specify that
the source output is on.
•
You need to specify the center frequency, span, and
reference power level for the analyzer.
Steps
Actions
Notes