Displaying Data
2
Keysight 1000B Series Oscilloscopes User’s Guide
61
2
In the Display menu, press
Type
to toggle the waveform display between:
To clear the display
1
Press
[Display]
.
2
In the Display menu, press
Clear
.
To set waveform persistence
1
Press
[Display]
.
2
In the Display menu, press
Persist
to toggle the waveform display between:
To adjust waveform intensity
1
Press
[Display]
.
2
In the Display menu, press
Intensity
and turn the entry knob to adjust the
waveform intensity.
To change the grid
1
Press
[Display]
.
2
In the Display menu, press
Grid
, and continue pressing the softkey or turn the
entry knob to select between:
Vectors
The oscilloscope connects the sample points by using digital
interpolation.
Digital interpolation maintains linearity by using a sin(x)/x
digital filter. The digital interpolation is suitable for real time
sampling and is most effective at 20 ns or faster horizontal
scale settings.
Dots
The sample points are displayed.
Infinite
Sample points remain displayed until the display is cleared or
persistence is set to “OFF”.
OFF
Содержание 1000B Series
Страница 1: ...Keysight 1000B Series Oscilloscopes User s Guide ...
Страница 12: ...12 Keysight 1000B Series Oscilloscopes User s Guide Contents ...
Страница 16: ...16 Keysight 1000B Series Oscilloscopes User s Guide Tables ...
Страница 36: ...36 Keysight 1000B Series Oscilloscopes User s Guide 1 Getting Started ...
Страница 53: ...Displaying Data 2 Keysight 1000B Series Oscilloscopes User s Guide 53 Figure 21 Waveform After Inversion ...
Страница 89: ...Capturing Data 3 Keysight 1000B Series Oscilloscopes User s Guide 89 Figure 37 Alternate Triggers ...
Страница 124: ...124 Keysight 1000B Series Oscilloscopes User s Guide 5 Saving Recalling and Printing Data ...
Страница 126: ...126 Keysight 1000B Series Oscilloscopes User s Guide 6 Oscilloscope Utility Settings Figure 53 Utility Key ...
Страница 138: ...138 Keysight 1000B Series Oscilloscopes User s Guide 6 Oscilloscope Utility Settings ...