Making Measurements
4
Keysight 1000B Series Oscilloscopes User’s Guide
105
Phase Between Rising Edges
Measures the phase between two waveforms using the rising edges.
Phase is the calculated phase shift from source 1 to source 2, expressed in
degrees. Negative phase shift values indicate that the rising edge of source 1
occurred after the rising edge of source 2.
Phase Between Falling Edges
Measures the phase between two waveforms using the falling edges. See
Figure 45
Phase Measurements
Phase
Delay
Source 1 Period
--------------------------------------
360
°
×
=
Phase from channel 1
to channel 2 rising edge
Time
origin
line
50%
Channel 2
Channel 1
50%
Delay
Period