Getting Started
1
Keysight 1000B Series Oscilloscopes User’s Guide
27
1
Using the BNC adapter, connect the probe to a square wave generator.
2
Set the square wave generator to a frequency of 1 MHz, an amplitude of
3 Vp-p, and an output termination of 50
Ω
.
3
Press the
[Auto Scale]
front panel key.
4
If waveform does not appear like the Correctly Compensated waveform shown
, then use a nonmetallic tool to adjust the 2 high frequency
compensation adjustments on the probe for the flattest square wave possible.
Figure 5
High Frequency Probe Compensation
High frequency compensation adjustments
Over Compensated
Under Compensated
Correctly Compensated