26
Keysight 1000B Series Oscilloscopes User’s Guide
1
Getting Started
Step 6. Compensate probes
Compensate probes to match your probe to the input channel. You should
compensate a probe whenever you attach it for the first time to any input channel.
Low Frequency Compensation
For the supplied passive probes:
1
Set the Probe menu attenuation to 10X. If you use the probe hooktip, ensure a
proper connection by firmly inserting the tip onto the probe.
2
Attach the probe tip to the probe compensation connector and the ground lead
to the probe compensator ground connector.
3
Press the
[Auto Scale]
front panel key.
4
If waveform does not appear like the Correctly Compensated waveform shown
, then use a nonmetallic tool to adjust the low frequency
compensation adjustment on the probe for the flattest square wave possible.
High Frequency Compensation
For the supplied passive probes:
Figure 4
Low Frequency Probe Compensation
Low frequency compensation adjustment
Over Compensated
Under Compensated
Correctly Compensated
Содержание 1000B Series
Страница 1: ...Keysight 1000B Series Oscilloscopes User s Guide ...
Страница 12: ...12 Keysight 1000B Series Oscilloscopes User s Guide Contents ...
Страница 16: ...16 Keysight 1000B Series Oscilloscopes User s Guide Tables ...
Страница 36: ...36 Keysight 1000B Series Oscilloscopes User s Guide 1 Getting Started ...
Страница 53: ...Displaying Data 2 Keysight 1000B Series Oscilloscopes User s Guide 53 Figure 21 Waveform After Inversion ...
Страница 89: ...Capturing Data 3 Keysight 1000B Series Oscilloscopes User s Guide 89 Figure 37 Alternate Triggers ...
Страница 124: ...124 Keysight 1000B Series Oscilloscopes User s Guide 5 Saving Recalling and Printing Data ...
Страница 126: ...126 Keysight 1000B Series Oscilloscopes User s Guide 6 Oscilloscope Utility Settings Figure 53 Utility Key ...
Страница 138: ...138 Keysight 1000B Series Oscilloscopes User s Guide 6 Oscilloscope Utility Settings ...