Keysight U3810 Series Getting Started Guide
74
Digital Pressure Sensor
Parameter
Minimum
Maximum
Typical
Units
Condition
VCC
2.0
3.6
3.3
Volts
Iin
265
40
µA
Relay Off
Vinh
.75
Volts
Vinl
.3
Volts
Calibrated
range
50
110
kPa
Noise
19
Pa rms
1x oversample
Absolute
accuracy
-.4
.4
+/-.4
kPa
Relative
accuracy
+/-.05
kPa
Relative accuracy during pressure
change between 70 to 110 kPa at any
constant temperature between −10 °C
to 50 °C
Resolution
.25 (.0625)
1.5(.3)
Pa(m)/LSB
Output data
rate
100
Hz
One-shot mode