U3050BH11
System Setup with N5245A
4. Remove the two upper feet from the U3050BH11 and install the two locking feet
N5242-20206 (right foot) and the N5242-20207 (left foot) using two screws (0515-2317) as
5. Place the PNA-X on top of the U3050BH11. Ensure that the front frame of the PNA-X is
positioned slightly forward of the locks that are attached to the Test Set. Slide the PNA-X
back so the locks engage the front frame of the analyzer as shown in
Figure
7
Rear Panel N5245A with U3050BH11
Figure 8
Locking the Analyzers
N5242-20206
N5242-20207
Slide the network analyzer
back to engage the Lock Link
in the back of the network
analyzer's front frame
Screw
Network Analyzer
Front Frame
Test Set
Front Frame
Lock Link
10
Keysight U3050BH11 User's and Service Guide
Содержание U3050BH11
Страница 4: ...Contents 2...
Страница 5: ...U3050BH11 Keysight U3050BH11 User s and Service Guide 1...
Страница 21: ...U3050BH11 Operational Check Figure 13 Keysight U3050BH11 User s and Service Guide 17...
Страница 23: ...U3050BH11 Operational Check Figure 15 Keysight U3050BH11 User s and Service Guide 19...
Страница 25: ...U3050BH11 Operational Check Figure 17 Keysight U3050BH11 User s and Service Guide 21...
Страница 27: ...U3050BH11 Operational Check Figure 19 Keysight U3050BH11 User s and Service Guide 23...
Страница 29: ...U3050BH11 Operational Check Figure 21 Keysight U3050BH11 User s and Service Guide 25...
Страница 31: ...U3050BH11 Operational Check Figure 23 Keysight U3050BH11 User s and Service Guide 27...
Страница 33: ...U3050BH11 Operational Check Figure 25 Keysight U3050BH11 User s and Service Guide 29...
Страница 37: ...U3050BH11 Theory of Operation Figure 27 U3050BH11 Block Diagram Keysight U3050BH11 User s and Service Guide 33...