U3050BH11
Troubleshooting
Front Panel LED is not Illuminated
If the U3050BH11 power supplies are operating the following procedures can be used to verify
the failure. Front Panel LED Assembly (U3050-6000
5
).
1.
If the LED is Off, verify the voltages (+
9
Volt) on the Power Supply. If the voltages are correct,
verify that the cables are connected correctly on both
the amplifier
and the LED assembly.
2.
Verify the amplifier switches are in the ON position and that the status light is
illuminated
.
Troubleshooting Diagrams and Tables
Table
9
Power Supply Connections
Power Supply
Terminals
Wire Color
−
9
V
Black
+12 V
White
−
5
V
Black
(
thin
)
+
5
V
Red (thin)
Keysight U3050BH11 User's and Service Guide
35
Содержание U3050BH11
Страница 4: ...Contents 2...
Страница 5: ...U3050BH11 Keysight U3050BH11 User s and Service Guide 1...
Страница 21: ...U3050BH11 Operational Check Figure 13 Keysight U3050BH11 User s and Service Guide 17...
Страница 23: ...U3050BH11 Operational Check Figure 15 Keysight U3050BH11 User s and Service Guide 19...
Страница 25: ...U3050BH11 Operational Check Figure 17 Keysight U3050BH11 User s and Service Guide 21...
Страница 27: ...U3050BH11 Operational Check Figure 19 Keysight U3050BH11 User s and Service Guide 23...
Страница 29: ...U3050BH11 Operational Check Figure 21 Keysight U3050BH11 User s and Service Guide 25...
Страница 31: ...U3050BH11 Operational Check Figure 23 Keysight U3050BH11 User s and Service Guide 27...
Страница 33: ...U3050BH11 Operational Check Figure 25 Keysight U3050BH11 User s and Service Guide 29...
Страница 37: ...U3050BH11 Theory of Operation Figure 27 U3050BH11 Block Diagram Keysight U3050BH11 User s and Service Guide 33...