U3050BH11
General Specifications
Maximum Power Levels
It is recommended that you do not operate components near damage levels
(+30 dBm). The power levels must be 3 dB below maximum level to ensure
Refer to your PNA-X specifications to determine the maximum input power
levels for the access and test ports, or to optimize the power levels in the
receivers.
Damage and maximum levels are not necessarily the optimum level.
Table
4
Test Set Power Levels & Front Panel Connections
Maximum Test Port RF Power Levels:
RCVR R1 IN
+15 dBm ±
7
Vdc
Interface Ports:
Port 1
+
30
dBm 0 Vdc
Port 1 CPLR THRU
+30 dBm 0 Vdc
Port 1 SOURCE OUT
+
26
dBm 0 Vdc
Port 1 RCVR R1 IN
+
15
dBm
7
Vdc
DUT OUT
1
1.
If CPLR THRU is connected to RCVR B IN on the
PNA-X
,
then
DUT OUT
max power level is
15 dBm 7 Vdc.
+
30 dBm
0
Vdc
+
15 dBm
0
Vdc
Port 2 CPLR ARM
+30 dBm
7
Vdc
Port 2 SOURCE OUT
+30 dBm 0 Vdc
Port 3
+2
6
dBm 0 Vdc
Access Ports:
PXA
+30 dBm 7 Vdc
Port
4
+
30
dBm 0 Vdc
6
Keysight U3050BH11 User's and Service Guide
Содержание U3050BH11
Страница 4: ...Contents 2...
Страница 5: ...U3050BH11 Keysight U3050BH11 User s and Service Guide 1...
Страница 21: ...U3050BH11 Operational Check Figure 13 Keysight U3050BH11 User s and Service Guide 17...
Страница 23: ...U3050BH11 Operational Check Figure 15 Keysight U3050BH11 User s and Service Guide 19...
Страница 25: ...U3050BH11 Operational Check Figure 17 Keysight U3050BH11 User s and Service Guide 21...
Страница 27: ...U3050BH11 Operational Check Figure 19 Keysight U3050BH11 User s and Service Guide 23...
Страница 29: ...U3050BH11 Operational Check Figure 21 Keysight U3050BH11 User s and Service Guide 25...
Страница 31: ...U3050BH11 Operational Check Figure 23 Keysight U3050BH11 User s and Service Guide 27...
Страница 33: ...U3050BH11 Operational Check Figure 25 Keysight U3050BH11 User s and Service Guide 29...
Страница 37: ...U3050BH11 Theory of Operation Figure 27 U3050BH11 Block Diagram Keysight U3050BH11 User s and Service Guide 33...