User’s and Service Guide U3047-90002
61
U3047AE04
Service Information
Receiver Signal Path Insertion Loss Test (S21)
Test Set Port 5, RCVR OUT to CPRL ARM
•
Set Test I/O addr.data 0.0
•
Connect PNA Port 1 flex cable to Test Set Port 5 CPLR ARM.
•
Connect PNA Port 2 flex cable to Test Set Port 5 RCVR OUT.
Test Set Port 6 RCVR OUT to CPRL ARM
•
Set Test I/O addr.data 16.0
•
Connect PNA Port 1 flex cable to Test Set Port 6 CPLR ARM.
•
Connect PNA Port 2 flex cable to Test Set Port 6 RCVR OUT.
Test Set Port 7, RCVR OUT to CPRL ARM
•
Set Test I/O addr.data 32.0
•
Connect PNA Port 1 flex cable to Test Set Port 7 CPLR ARM.
•
Connect PNA Port 2 flex cable to Test Set Port 7 RCVR OUT.
Test Set Port 8, RCVR OUT to CPRL ARM
•
Set Test I/O addr.data 64.0
•
Connect PNA Port 1 flex cable to Test Set Port 8 CPLR ARM.
•
Connect PNA Port 2 flex cable to Test Set Port 8 RCVR OUT.
Test Set Port 5, RCVR OUT to Port 5
•
Set Test I/O addr.data 0.16
•
Connect PNA Port 1 flex cable to Test Set Port 5.
•
Connect PNA Port 2 flex cable to Test Set Port 5 RCVR OUT.
Test Set Port 6, RCVR OUT to Port 6
•
Set Test I/O addr.data 16.16
•
Connect PNA Port 1 flex cable to Test Set Port 6.
•
Connect PNA Port 2 flex cable to Test Set Port 6 RCVR OUT.
Test Set Port 7, RCVR OUT to Port 7
•
Set Test I/O addr.data 32.16
•
Connect PNA Port1 flex cable to Test Set Port 7.
•
Connect PNA Port 2 flex cable to Test Set Port 7 RCVR OUT.
Test Set Port 8, RCVR OUT to Port 8
•
Set Test I/O addr.data 64.16
•
Connect PNA Port 1 flex cable to Test Set Port 8.
•
Connect PNA Port 2 flex cable to Test Set Port 8 RCVR OUT.