10
User’s and Service Guide U3047-90002
U3047AE04
Front and Rear Panel Features
Figure 3
U3047AE04 Rear Panel
Always use the three-prong
AC
power cord supplied with this
product. Failure to ensure adequate grounding by not using this cord
may cause damage to the product.
Control Lines/Voltage Adjust
Line Module Test Set Interface Pass Through Interface Chassis Ground
Control Lines and Voltage Adjust
For further information pertaining to control lines and voltage adjustments see. Refer to
Chassis Ground
A threaded terminal post for connecting the Test Set to a conductive object, cabinet or
structure to ensure a common potential and reduce leakage current in a system. Requires an
English 1/4-20 thread nut (2950-0004) and lock washer (2190-0067).
Pass Through Interface
The Pass Through Interface is used to connection to another test set.
Test Set Interface
The Test Set Interface connector is used to send address and data to the Test Set from the
PNA-X.
Line Module
The line module contains the power cord receptacle.The line fuse, as well as a spare, reside
within the line module.
illustrates where the fuses are located and how to access them.
Install the instrument so that the detachable power cord is readily identifiable and is easily
reached by the operator. The detachable power cord is the instrument disconnecting device. It
disconnects the mains circuits from the mains supply before other parts of the instrument. The
standby switch is only a standby switch and is not a LINE switch. Alternatively, an externally
installed switch or circuit breaker (which is readily identifiable and is easily reached by the
operator) may be used as a disconnecting device.