User’s and Service Guide U3047-90002
U3047AE04
Cal Kit Operational Check
7.
for Ports 2 thru 8. When finished, there should be 8 Cal Sets
saved with the titles “999.1” thru “999.8” (8-Port).
8. Select Trace/Chan > Trace > Delete Trace. There should be no traces on the PNA-X screen.
9. To launch the Cal Set Viewer toolbar. Select Response > Cal > Manage Cals >
Cal Set Viewer.
Figure 36 Calibration, Cal Set Viewer
10.
From the Cal Sets drop-down menu, select 999.1 and select Enable. Select the Reflection
Tracking(x,x), where x,x is the port being tested, term in the center
drop-down menu and
ensure that the Enable and Error Terms check boxes are selected. See
11.
Compare the Reflection Tracking (1,1) trace to the appropriate limits in
The trace should be above the limit.
12.