User’s and Service Guide U3047-90002
59
U3047AE04
Service Information
RF Switch Path Testing
If you suspect an RF signal path problem, the following procedure will confirm all RF signal
paths through the Test Set.
Equipment Required
•
N5247A PNA-X
•
Two RF Flex Cables with 1.85 mm connectors (8121-1445 or equivalent)
•
Three 1.85 mm Adapters (female to female)
•
Test Set I/O Cable (8120-6818)
Equipment Setup
1. Connect the Test Set I/O cable (8120-6818) from the PNA to Test Set. Refer to
. Set PNA to “Standalone mode” (Test can not be performed in Multiport mode).
2. Connect the RF flex cables with the adapters to the PNA Ports 1 and 2. Connect the cables
together using a 1.85 mm adapter.
3. Set the following:
Table 1
6
RF Signal Path Insertion Loss (S21)
RF Path Description
Signal
Insertion
Loss
(typical)
Source IN to CPLR THUR
Source
Source IN to Ports 5 - 8
Source
RCVR OUT to CPLR ARM
Receiver
RCVR Out to Ports 5 - 8
Receiver
•
Frequency range: 10 MHz to 70 GHz, (or the frequency range of the PNA you are using)
•
Ref Level: 0 dB
•
Scale: 10 dB
•
Power Level: 0 dBm
•
IF BW: 1 kHz
4. Select measure S21 trace and perform a Trace Normalization.