PNA-L RF Interface Cable Connections
illustrates the cable configuration of the test set to the PNA-L.
The cables have been supplied with the kit (U3021-60045).
1.
Remove the SOURCE OUT to CPLR THRU and RCVR IN to CPLR ARM jumpers
(x16) on the PNA-L. The RCVR R1 to SOURCE OUT reference loop jumper (x1)
remain
on the front panel.
2.
Connect the RF interconnect cables, supplied with this option, from the PNA-L to
the
test set. Torque each cable to 8 in-lb. The Z5623-20418 are the short cables
and
the
Z5623-20419 are the long cables. Refer to
below
Each end of the interconnect RF cables have a different length from the bend. When
connecting the RF Interconnect cables be sure that the longer end (from the bend) is
connected to the PNA-L.
Over torque will cause damage to the test set and may cause the connectors to spin
or become loose.
Table
8
PNA-L Interface Cable Connection (U3021-60045)
RF Cables
From PNA-L
To Test Set
Z5623-20418
SOURCE OUT
SOURCE OUT
Z5623-20418
CPLR ARM
CPLR ARM
Z5623-20419
CPLR THRU
CPLR THRU
Z5623-20419
RCVR IN
RCVR OUT
22
U3042AE12 User's and Service Guide