Setting the Control Lines with Address and Data Values
This section describes how to control the rear panel DUT control lines.
The following control feature will function only while the analyzer is in Standalone
Mode.
Setting the PNA to Standalone Mode
1.
Select
Utility
>
System
>
Configure
>
Multiport Capability
-OR-
(Instrument > Setup > External Hardware > Multiport > Multiport Configuration...)
.
In the dialog box
, select
Restart as a standalone PNA > OK
.
2.
Access Interface Control Mode
As described in
“Setting the Test Port Paths with Address and Data” on page
, the
<address>.<data> values are determined in a similar
manner
, with the following
exceptions:
•
• The <address> value is always = 112
Table 1
5
Test Set DUT Control Address and Data Logic Table
<Address>
<Data>
Description
Line/Pin
112
0
ALL DUT Control Lines set to logic high or connected to Pin 13
1-8
112
255
ALL DUT Control Lines set to logic high or connected to Pin 14
1-8
112
1
DUT Control Line 1 set to logic low or connected to Pin 14
1
112
2
DUT Control Line 2 set to logic low or connected to Pin 14
2
112
4
DUT Control Line 3 set to logic low or connected to Pin 14
3
112
8
DUT Control Line 4 set to logic low or connected to Pin 14
4
112
16
DUT Control Line 5 set to logic low or connected to Pin 14
5
112
32
DUT Control Line 6 set to logic low or connected to Pin 14
6
112
64
DUT Control Line 7 set to logic low or connected to Pin 14
7
112
128
DUT Control Line 8 set to logic low or connected to Pin 14
8
58
U3042AE12 User's and Service Guide