3. Connect the RF Flex cables to the Source IN and CPLR THRU port
as
indicated in
. T
he expected results should be similar to
shown below
.
Figure 6
6
Source IN to CPLR THRU & RCVR OUT to CPLR ARM Path Response
Table 2
4
Source Bypass Signal Path Connections and Commands
Path #
Test Port Group
RF Path Description
Control Mode
<Address>.<Data>
Path Components
1
Port 5, 9, 13
Source IN to CPLR THRU
0.1
S100
2
Port 6, 10, 14
Source IN to CPLR THRU
16.1
S200
3
Port 7, 11, 15
Source IN to CPLR THRU
32.1
S300
4
Port 8, 12, 16
Source IN to CPLR THRU
64.1
S400
84