Zero Set (Peak Path)
Description
Zero set is defined as the amount of residual offset error that is present following a
zero operation. This offset error is caused by contamination from several sources,
including circuit noise. This test measures the effectiveness of zero set by
performing 15 back-to-back zero operations of the peak path (with no sensor
attached), after which the standard deviation of the results is calculated and
returned as the measured value.
Equipment
– No test equipment required
Test method
1
Execute the internal zero set measurement procedure for Channel A.
2
Read back the result of the measurement from the DUT.
3
Repeat this procedure for Channel B.
4
The test will take a few minutes to complete.
5
The measurement result should be less than 0.015. The smaller the
measurement result, the smaller the amount of residual offset error.
NOTE
– This test can be performed manually via the commands:
SERV:BIST:PEAK[1|2]:ZSET
SERV:BIST:PEAK[1|2]:ZSET:NUM?
(Refer to the
Programming Guide
for further details on the use of these
commands)
– No adjustment is available for this test if it fails (see
Содержание N8262A
Страница 1: ...Keysight N8262A P Series Modular Power Meter Service Guide ...
Страница 18: ...18 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 20: ...20 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...22 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 40: ...1 Specifications 40 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 58: ...2 Performance Tests 58 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 62: ...3 Adjustments 62 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 70: ...4 Theory of Operation 70 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...