Linearity (Average Path)
Description
Linearity over the full input voltage range of the measurement path is warranted
to provide assurance of instrument accuracy. This test measures Linearity by
using a calibration DAC and a calibration ADC (built into the DUT) to stimulate
and compare performance of the Average Path against the measurement ADC,
returning the worst case percentage error.
Equipment
– No test equipment required
Test method
1
Execute the internal linearity measurement procedure for Channel A.
2
Read back the result of the measurement from the DUT.
3
Repeat this procedure for Channel B.
4
The test will take a few minutes to complete.
5
The measurement result should be less than
0.5 and greater than –0.5. The
optimum measurement result for this test is 0.
NOTE
– This test can be performed manually via the commands:
SERV:BIST:CW[1|2]:LIN
0
SERV:BIST:CW[1|2]:LIN:PERR?
(Refer to the
Programming Guide
for further details on the use of these
commands)
– No adjustment is available for this test if it fails (see
Содержание N8262A
Страница 1: ...Keysight N8262A P Series Modular Power Meter Service Guide ...
Страница 18: ...18 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 20: ...20 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...22 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 40: ...1 Specifications 40 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 58: ...2 Performance Tests 58 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 62: ...3 Adjustments 62 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 70: ...4 Theory of Operation 70 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...