Performance Tests
2
Keysight N8262A Service Guide
53
Figure 2-1
Absolute accuracy test setup
Test method
1
Connect the equipment as shown in
2
Unplug the power cord from the range calibrator. Eliminate ground loops to
enable the 11683A (Option H01) range calibrator to operate properly.
11683A range
calibrator front panel
11683A range
calibrator rear panel
N8262A P-Series
modular power meter
BNC T-joint connector
3458A digital
multimeter
BNC cable
BNC cable
Banana
connector
Banana
connector
33250A function
generator
NOTE
For the Keysight N8262A P-Series modular power meter, the equivalent key
presses should be performed on both channels.
Содержание N8262A
Страница 1: ...Keysight N8262A P Series Modular Power Meter Service Guide ...
Страница 18: ...18 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 20: ...20 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...22 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 40: ...1 Specifications 40 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...