Troubleshooting Guide
5
Keysight N8262A Service Guide
77
Performance Test
Type of failures
Debug tips
Possible faults
1 mW power reference level
failures
Attempt to adjust the 1 mW
Power Reference Level
Calibrator Assembly (high
probability)
Mother Board (low probability)
VSWR failures
Not applicable
Calibrator Assembly
Time base frequency accuracy
failures
– Check that the SMB cable
being used is not damaged
– Check that the SMB is
connected to ‘Trig Out’, not
‘Trig In’
Mother Board
Zero set (average path) failures
Not applicable
Mother Board
Zero set (peak path) failures
Not applicable
Mother Board
Linearity (average path) failures
Not applicable
Mother Board
Linearity (peak path) failures
Replace the measurement board
assembly for the channel to see if
this clears the fault
– Measurement board assembly
(low probability)
– Mother board (high
probability)
Rise/Fall time (peak path) failures Check/reseat the sensor flex RF
connections
– Sensor flex assembly (low
probability)
– Measurement board assembly
(low probability)
– Mother board (high
probability)
Содержание N8262A
Страница 1: ...Keysight N8262A P Series Modular Power Meter Service Guide ...
Страница 18: ...18 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 20: ...20 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 22: ...22 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 40: ...1 Specifications 40 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 58: ...2 Performance Tests 58 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 62: ...3 Adjustments 62 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Страница 70: ...4 Theory of Operation 70 Keysight N8262A Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...