6
Repair Guide
116
Keysight N8262A Service Guide
Replacing the PPMC Assembly
– The PPMC Assembly is pre-programmed with N8262A firmware.
– Always perform a firmware update to the instrument if the PPMC assembly has
been replaced.
– Instrument serial number:
This can be stored in the PPMC assembly via the command:
SERV:
SNUM
<CHARACTER
DATA>
– Instrument option(s):
This/these can be stored in the PPMC assembly via the command:
SERV:
OPT
“<CHARACTER
DATA>”
Refer to the
Programming Guide
for further details on the use of these commands.