14 Keysight N5280A User’s and Service Guide
N5280A
General Specifications
-
Dimensions and Space Requirements
Standard installation of the N5280A and PNA includes configuration and
installation on a customer provided lab bench, or table top of adequate size
and strength.
Maximum Power Levels
Table 5
Instrument Dimensions
Model
Weight
Height
Wid th
Depth
N5242A 37 kg
82 lb (±0.5 lb)
26.67 cm
(10.5 in)
42.5 cm
(16.75 in)
55.8 cm
(21.97 in)
N5264A 22 kg
48 lb (±0.5 lb)
26.67 cm
(10.5 in)
42.5 cm
(16.75 in)
55.8 cm
(21.97 in)
N5280A
9.1 kg
(20 lb)
8.9 cm
(3.5 in)
42.5 cm
(16.75 in)
48.3 cm
(19 in)
Table 6
Power Levels
RF Input Power Damage Levels
RF Port
+18 dBm
LO Port
+5 dBm
Optimum LO Power
0 dBm (±1 dB)
RF Input @ 0.1 dB Typical Compression
A- D Receiver
10 dBm
IF Output Level
1
1.
IF Output level is based on the RF Input @
0.1 dB typical compression.
Ports A- D
10 dBm
Refer to your PNA specifications to determine the maximum input power
levels for the PNA access and test ports, or to optimize the power levels in
the receivers.
Damage and maximum levels are not necessarily the optimum level.