12 Keysight N5280A User’s and Service Guide
N5280A
General Specifications
-
Power Requirements
Verify that the required ac power is available before installing the Test Set to
the PNA.
— 100/120/220/240 VAC (50/60Hz)
— The instruments can operate with mains supply voltage fluctuations up to ±
10% of the nominal voltage.
— Air conditioning equipment (or other motor–operated equipment) should
not be placed on the same ac line that powers the Test Set and PNA.
—
Table 3
contains the maximum wattage for all instruments. This table can
be use to determine the electrical and cooling requirements.
Table 3
Maximum Power Requirements
Standard Equipment
Instrument
Max Wattage
N5242A
450
N5264A
450
N5280A
350
This is a Safety Class I product (provided with a protective earthing
ground incorporated in the power cord). The mains plug shall be
inserted only into a socket outlet provided with a protective earth
contact. Any interruption of the protective conductor, inside or
outside the instrument, is likely to make the instrument dangerous.
Intentional interruption is prohibited.