42 Keysight N5280A User’s and Service Guide
N5280A
Operational Check
-
Check LO AUX
1.
Connect the power meter to the LO AUX on the N5280A.
2.
Use the output frequencies and power levels from the
Table 15 on
page 40
. Record the result in table for the measure LO AUX. The LO AUX
specification is greater than
10 dBm.
Figure 28
IF LO AUX Setup
IF OUTPUTS
N5280A
Front Panel
LO Source
RF INPUTS
A
B
C
D
B
C
D
LO INPUT
LO IN
LO AUX
D/R2
C/R1
A
B
IF OUTPUTS
A
B
C
D
ATTENUATOR
IF OUTPUTS
N5280A
Rear Panel
A
Power Sensor
Power Meter