Making Measurements
Keysight
N4374B Lightwave Component Analyzer, User’s Guide
91
Please also refer to the documentation supplied with the
patchcord.
a For the user calibration select an electrical calibration as
made according to the procedure given in “Electrical
calibration for single-ended measurements” on page 75.
This calibration must have the same network analyzer
settings as you are using to measure your DUT.
This will only be used for the “user calibration”
measurements. Afterwards the LCA returns to the
original electrical calibration for the DUT measurements.
The currently used calibration is preselected.
b Select “Reduce IFBW for user calibration measurement” to
reduce noise on the user calibration data. The default
bandwidth reduction factor is set in the file
LCAConfig.xml.
c Click
to proceed
.
d When asked whether you want to save the data
• If you can use this calibration data again, click
,
give your calibration data file a name, and save it.
If you want to save the complete LCA setup, make sure to
select “Also save the LCA Settings” . See “Saving and
Loading Setups” on page 61 for more information about
LCA setups.
• If you will have no need to reuse or refer to this
calibration data, click
.
Содержание N4374B
Страница 1: ...Keysight N4374B Lightwave Component Analyzer User s Guide ...
Страница 2: ......
Страница 3: ......
Страница 22: ......
Страница 42: ......
Страница 58: ......
Страница 73: ...User Interface Reference Keysight N4374B Lightwave Component Analyzer User s Guide 71 ...
Страница 74: ......
Страница 96: ......
Страница 114: ......
Страница 118: ......
Страница 124: ......
Страница 125: ...Definition of Terms 10 Definition of Terms Definition of Terms 124 ...
Страница 156: ......
Страница 160: ......
Страница 163: ......