Definition of Terms
130
Keysight
N4374B Lightwave Component Analyzer, User’s Guide
Measurement:
Based on a reference RF O/E receiver.
Responsivity
The amplitude response
R
of a device under test (DUT) to a
harmonic stimulus:
• E/O measurement:
R
is the ratio of
in Watt to the electrical stimulus amplitude in Ampere,
expressed in W/A.
• O/E measurement:
R
is the ratio of electrical response
in Watt,
expressed in A/W.
In both cases, responsivity in
(“dB
W/A
“
or “dB
A/W
“) is calculated as
20
-log(
R
).
• O/O measurement:
R
is the difference between the
in dBm at the DUT output and at the DUT
input, expressed in dBo. If expressed in dBe, it is twice this
number.
References
(1)
“Guide to the Expression of Uncertainty in Measurement”
(“GUM”), BIPM, IEC, ISO et al. (1993)
N O TE
Changes in the cabling after user e-cal add to the specified relative
frequency response uncertainty.
N O TE
Responsivity of an O/E or E/O device may be called
conversion efficiency
.
For an O/O device,
R
[dBo] = -(insertion loss [dB]).
Содержание N4374B
Страница 1: ...Keysight N4374B Lightwave Component Analyzer User s Guide ...
Страница 2: ......
Страница 3: ......
Страница 22: ......
Страница 42: ......
Страница 58: ......
Страница 73: ...User Interface Reference Keysight N4374B Lightwave Component Analyzer User s Guide 71 ...
Страница 74: ......
Страница 96: ......
Страница 114: ......
Страница 118: ......
Страница 124: ......
Страница 125: ...Definition of Terms 10 Definition of Terms Definition of Terms 124 ...
Страница 156: ......
Страница 160: ......
Страница 163: ......