Specifications
Keysight
N4374B Lightwave Component Analyzer, User’s Guide
133
• Number of averages: 1
• IFBW 300 Hz
• Network analyzer set to “stepped sweep – sweep moves in
discrete steps”
• After full two-port electrical calibration using an Electronic
Calibration Module,
Keysight
85092C, at constant temperature
(±1° C)
• Modulator bias optimization set to “every sweep”
• Measurement frequency grid equals electrical calibration
grid
• DUT signal delay = 0.1/IF-BW
• Specified temperature range: +20° C to +26° C.
• After warm-up time of 90 minutes
• Using high quality electrical and optical connectors and RF
cables in perfect condition
• Using supplied RF cables (8120-8862)