Performance Quick Check
Keysight
N4374B Lightwave Component Analyzer, User’s Guide
35
• Laser power: +4 dBm
• Optical input: Input 1
• User Calibration: None
7. Click
, then
to set the output power.
8 Press
to start optical power meter application.
9 Read the optical power from the power meter application.
The readout must be +4 dBm ± 0. 5dB
10.Press the [Macro Setup] key on the front panel of the
network analyszer, then select
, to see
.
11.Press
to start the LCA measurement setup.
12.Select the OO measurement tab.
13.Set the following parameters:
• If you have option 102 (Dual wavelength 1310nm,
1550nm), set Wavelength to the wavelength at which you
are testing your DUTs (1310nm or 1550nm).
• Laser power: -2 dBm
• Optical input: input1 (“+7 dBm max”) or input 2
(“+17 dBm max”).
• User Calibration: None
14.Click
, then
to set the output power.
15 Press
to start optical power meter application
16.Read the optical power from the power meter application.