Measurement Concepts
Keysight
N4374B Lightwave Component Analyzer, User’s Guide
51
The calibration is with respect to the optical connectors on the
front of the optical test set. Whenever there are additional
optical adapters or elements connecting to the device under
test, these optical paths can be accounted for by "optical path
deeembedding" (see “Optical Path Deembedding” on page 66).
The systematic responses of these two stages of calibration can
then be deembedded, yielding the response of the device under
test (DUT). (See “Signal Relationships in Opto-electric Devices”
on page 46 for more detail.)