Performance Quick Check
36
Keysight
N4374B Lightwave Component Analyzer, User’s Guide
The readout must be -2 dBm ± 0. 5dB
If the results differ from the expected
• Try cleaning the optical connections as described in
• Switch the laser off and inspect all optical connectors with a
microscope.
• If the connectors on the N4374B Lightwave Component
Analyzer seem to be damaged contact Keysight for repair.
• If the connectors on the patchcord seem to be damaged
contact Keysight for replacements.
• With an external optical power meter, you can verify the
output power of the internal laser and the optical connection
on the transmitter side.
If the problem persists contact Keysight for support.