30
6000 X-Series Oscilloscopes Service Guide
2
Testing Performance
2
Set up the oscilloscope.
a
Adjust the horizontal scale to
200.0 us/div
.
b
Set the Volts/Div setting to the value in the first line in
(depending
on the oscilloscope model).
c
Adjust the channel’s vertical position knob to place the baseline (reference
level) at 0.5 major division from the bottom of the display.
d
Press the
[Acquire]
key.
e
Then press the
Acq Mode
softkey and select
Averaging
.
f
Then press the
#Avgs
softkey and set it to 64.
Wait a few seconds for the measurement to settle.
Table 8
Settings Used to Verify DC Vertical Gain Accuracy
Volts/Div Setting
Precision Source Setting
Test Limits
5 V/Div
35 V
34 V
to
36 V
2 V/Div
14 V
13.6 V
to
14.4 V
1 V/Div
7 V
6.8 V
to
7.2 V
500 mV/Div
3.5 V
3.4 V
to
3.6 V
200 mV/Div
1.4 V
1.36 V
to
1.44 V
100 mV/Div
700 mV
680 mV
to
720 mV
50 mV/Div
350 mV
340 mV
to
360 mV
20 mV/Div
140 mV
136 mV
to
144 mV
10 mV/Div
70 mV
68 mV
to
72 mV
5 mV/Div
1
35 mV
34 mV
to
36 mV
2 mV/Div
1, 2
14 mV
13.6 mV
to
14.4 mV
1 mV/Div
1, 2
7 mV
6.6 mV
to
7.4 mV
1
A blocking capacitor is required at this range to reduce noise. See
2
Full scale is defined as 16 mV on the 1 mV/div range. Full scale on all other ranges is defined as 8
divisions times the V/div setting.
Содержание InfiniiVision 6000 X Series
Страница 1: ...Keysight InfiniiVision 6000 X Series Oscilloscopes Service Guide...
Страница 12: ...12 6000 X Series Oscilloscopes Service Guide Tables...
Страница 58: ...58 6000 X Series Oscilloscopes Service Guide 3 Calibrating and Adjusting...
Страница 82: ...82 6000 X Series Oscilloscopes Service Guide 4 Troubleshooting...