Testing Performance
2
6000 X-Series Oscilloscopes Service Guide
33
The difference in average voltage readings should be within the test limits of
(depending on the oscilloscope model).
If a result is not within the test limits, go to the “Troubleshooting” chapter. Then
return here.
11
Disconnect the precision source from the oscilloscope.
12
Repeat this procedure to check the DC vertical gain accuracy with the
remaining Volts/div setting values in
(depending on the oscilloscope
model).
13
Finally, repeat this procedure for the remaining channels to be tested.
Use a Blocking Capacitor to Reduce Noise
On the more sensitive ranges, such as 1 mV/div, 2 mV/div, and 5 mV/div, noise
may be a factor. To eliminate the noise, add a BNC Tee, blocking capacitor, and
shorting cap at the oscilloscope channel input to shunt the noise to ground. See
. If a BNC capacitor is not available, use an SMA blocking capacitor,
adapter, and cap. See
in the equipment list on
for
details.
Figure 4
Using a Blocking Capacitor to Reduce Noise
To oscilloscope input
BNC shorting
cap
Blocking
Capacitor
Содержание InfiniiVision 6000 X Series
Страница 1: ...Keysight InfiniiVision 6000 X Series Oscilloscopes Service Guide...
Страница 12: ...12 6000 X Series Oscilloscopes Service Guide Tables...
Страница 58: ...58 6000 X Series Oscilloscopes Service Guide 3 Calibrating and Adjusting...
Страница 82: ...82 6000 X Series Oscilloscopes Service Guide 4 Troubleshooting...